Skip to content
    • Contact
    • (02) 97486988
    • Contact
    • (02) 97486988
Domo-Technica AustraliaDomo-Technica Australia
  • About Us
  • Contact Us
  • 0
    Cart

    No products in the cart.

    Return to shop

  • Home
  • Fluke Biomedical
    • Defibrillator Analysers
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Neonatal Test Equipment
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Digital Pressure Meters
    • Patient Simulators
    • Phototherapy Radiometer
    • Rigid Endoscope Tester
    • Test Automation
  • Raysafe
    • X-Ray QA Instruments
    • Phantoms and Test Tools
    • Radiation Safety
    • Real-Time Personal Dosimeter
  • Quantum Design
    • Material Science
    • Cryogenics
    • Optics
    • Nanoscience
    • Sample Synthesis
    • Biotechnology & Chemistry
    • Magnetic Separator
    • Laboratory Squids
  • Montana Instruments
    • Research Platforms
    • Measurement Systems
    • Integrated Solutions
    • Accessories
  • Calibration
    • Australia
    • New Zealand
  • All Products
Add to Wishlist
Impulse 6000D Defibrillator Analyzer
Home / Fluke Biomedical / Defibrillator Analysers

Impulse 6000D Defibrillator Analyzer

The Impulse 6000D model is a Defibrillator Analyzer is a portable, rugged, easy-to-carry device with an intuitive user interface, backlight, and easy-to-read display.

Categories: Defibrillator Analysers, Fluke Biomedical

Browse
  • Fluke Biomedical
    • Defibrillator Analysers
    • Digital Pressure Meters
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Neonatal Test Equipment
    • Patient Simulators
    • Phototherapy Radiometer
    • Rigid Endoscope Tester
    • Test Automation
  • Montana Instruments
    • Accessories
    • Integrated Solutions
    • Measurement Systems
    • Research Platforms
  • Quantum Design
    • Biotechnology & Chemistry
    • Cryogenics
    • Laboratory Squids
    • Magnetic Separator
    • Material Science
    • Nanoscience
    • Optics
    • Sample Synthesis
    • SCL-Sensor.Tech – Self-Sensing Cantilevers
  • Raysafe
    • Phantoms and Test Tools
    • Radiation Safety
    • Real-Time Personal Dosimeter
    • X-Ray QA Instruments
  • Uncategorised
  • Description
  • Features
  • Specifications
  • Models and Accessories

Impulse 6000D Defibrillator Analyzer

Fluke Biomedical defibrillator analyzers ensure proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment, including mono- and bi-phasic defibrillators and AEDs. These best-in-class defibrillators offer test capabilities that encompass a global spectrum of established pulse shapes, feature breakthrough AED technology compatibility, and deliver superior performance with accuracy readings ± 1% of reading +0.1J.

The Impulse 6000D model is a Defibrillator Analyzer is a portable, rugged, easy-to-carry device with an intuitive user interface, backlight, and easy-to-read display. With functionality that includes energy output measurement and ECG wave testing, the Impulse 6000D features 10 independent ECG outputs that provide 12 lead combinations for standardized clinical signals.

Ansur test automation may be purchased in order to standardize test procedures, capture waveforms and test results, and evolve with increasing requirements for electronic data storage.

*Please Note:
The Impulse 6000D is not compatible with the 7010 Load Accessory.

 

 

 

Features

IEC 60601-2-4 Compliance
Lown, Edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility
AED compatibility
12-lead ECG simulation
Internal pacer brand selections
Flexible heart-rate settings (1BPM step)
DSP-based measurements for firmware and waveform upgrade
Waveform capture, store and replay
First-in-class accuracy ± 1% of reading + 0.1J
Long lasting, rechargeable battery

Specifications

General Specifications
Temperature, Operating  10 °C to 40 °C (50 °F to 104 °F)
Temperature, Storage  -20 °C to +60 °C (-4 °F to +140 °F)
Humidity  10 % to 90 % non-condensing
Display  LCD display
Communications  USB device port for computer control
Modes of Operation  Manual and remote
Power  Internal rechargeable NiMH battery pack for nine hours (typical) operation after full charge or the battery charger can operate the analyzer and charge the battery simultaneously
Battery Charger  100 V to 240 V input, 15 V/1.5 V A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle
Enclosure  ABS plastic housing
Dimensions  32 cm x 24 cm x 13 cm
 (13 in x 9.5 in x 5 in)
Weight  3.02 kg (6.6 lb, 0.1 oz)
Safety Standards  CE: IEC/EN61010-1 2nd Edition; Pollution degree 2 CAN/CSA-C22.2 No. 61010-1; UL61010-1 C-Tick: Australian EMC

 

Energy Output Measurement Specifications
Compatible Defibrillator Waveshapes
 Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
 AC Pulsed Biphasic waveform has not been approved in the United States.
Auto-ranged Measurement
 0.1 J to 600 J
Accuracy
 ± (1 % of reading + 0.1 J)
Load Resistance
Resistance  50 Ω
Accuracy  ±1 %, non-inductive (<2 µH)
Pulse Trigger Level
 20 V
Pulse Width
Range  1 ms to 50 ms
Accuracy  ±0.1 ms
Voltage
Range  20 V to 5000 V
Accuracy  ± (1 % of reading + 2 V)
Current
Range  0.4 A to 100 A
Accuracy  ± (1 % of reading + 0.1 A)
Tilt (Biphasic and Pulsed Biphasic)
Range  1 % to 99 %
Accuracy  ± 1 digit
Interphase Delay (Biphasic and Pulsed Biphasic)
Range  0.1 ms to 9.9 ms
Accuracy  ± 0.1 ms
Frequency (Pulsed Biphasic Only)
Range  2000 Hz to 8000 Hz
Accuracy  ± 1 % of reading
Duty Cycle (Pulsed Biphasic Only)
Range  1 % to 99 %
Accuracy  ± digit
Sample Rate
 250 kHz (4 µs sample)
Maximum Average Power
 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Scope Output
 Autorange:2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
 

  • Output: BNC
  • Output impedance: 50 Ω
  • Amplitude Accuracy: ± 5 %
Charge Time Measurement
Range  0.1 s to 100 s
Accuracy  ± 0.05 s, typical
Synchronization Test (Elective Cardioversion)
Delay Time Measurement  Timing window: : ECG R-wave peak to the defib pulse peak
 Range: -120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380 ms following the R-wave peak
 Resolution: 1 ms
 Accuracy: ± 1 ms
ECG waves  Normal Sinus Rhythm (NSR): 10 BPM to 180 BPM, in 1 BPM steps
 Atrial fibrillation: Coarse and fine
 Monomorphic Ventricular Tachycardia: 120 BPM to 240 BPM, in 5 BPM steps
 Asystole: Flat line
Automated Defibrillator Test ECG Waves
Normal Sinus  10 BPM to 300 BPM, in 1 BPM steps
Ventricular Fibrillation  Coarse and fine
Monomorphic Ventricular Tachycardia  120 BPM to 300 BPM, in 1 BPM steps
Polymorphic Ventricular Tachycardia  5 types
Asystole  Flat line

 

ECG Waves
ECG General
Lead configuration  12-lead simulation. RA, LL, LA, RL, V1-6 with independent signal outputs
Lead-to-lead impedance  1000 Ω (nominal)
Rate accuracy  ± 1 % of nominal
ECG Amplitudes
Reference lead  Selectable, Lead II (default) or Lead 1
Settings  0.05 mV to 0.45 mV, in 0.05 mV steps
0.5 mV to 5 mV, in 0.5 mV steps
Accuracy (All Performance waves and Normal Sinus R waves)  Lead II: ± 2 %
All other leads: ± 5 %
Defib paddles: ± 5 %
Amplitude of ECG signals relative to amplitude setting (in percent)  Lead II reference:
 Performance waves and R wave detection
 Lead I – 70 %
 Lead II – 100 %
 Lead III – 30 %
 Lead V1 – 100%
 Lead V2 – 100%
 Lead V3 – 100 %
 Lead V4 – 100 %
 Lead V5 – 100 %
 Lead V6 – 100 %
 Normal Sinus waves
 Lead I – 70 %
 Lead II – 100 %
 Lead III – 30 %
 Lead V1 – 24%
 Lead V2 – 48%
 Lead V3 – 100 %
 Lead V4 – 120 %
 Lead V5 – 112 %
 Lead V6 – 80 %
 Lead I reference:
 Performance waves and R wave detection
 Lead I – 100 %
 Lead II – 150 %
 Lead III – 50 %
 Lead V1 – 100%
 Lead V2 – 100%
 Lead V3 – 100 %
 Lead V4 – 100 %
 Lead V5 – 100 %
 Lead V6 – 100 %
 Normal Sinus waves
 Lead I – 100 %
 Lead II – 150 %
 Lead III – 50 %
 Lead V1 – 24%
 Lead V2 – 48%
 Lead V3 – 100 %
 Lead V4 – 120 %
 Lead V5 – 112 %
 Lead V6 – 80 %
ECG Normal Sinus
Rates  10 BPM to 360 BPM, in 1 BPM steps
ECG High Level Output (BNC Jack)
Amplitude Range  0.5 V/mV of reference lead setting
Accuracy  ± 5 %
Output Impedance:  50 Ω
ECG On Cefibrillator Input Load
 Same as the Lead II amplitude but limited to ± 4 mV
ECG Performance Waves
Square wave  2 Hz and 0.125 Hz
Triangular wave  2 Hz and 2.5 Hz
Sine waves  0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz
Pulse  30BPM and 60 BPM, 60 ms pulse width
R-Wave Detection
Waveform  Haver-triangle
Rates  30, 60, 80, 120, 200 and 250 BPM
Widths  8, 10, 12 and 20 to 200 (by 10) ms
Amplitudes  0.05 mV to 0.45 mV 0.05 mV hard return
 0.5 mV to 5 mV, in 0.5 mV steps
Accuracy  ± (1 % setting + 1 ms)
Noise Immunity
Wave  Sine
Line Frequency  50 Hz or 60 Hz ( ± 0.5 Hz)
Amplitude  Range: 0.0 mV to 10 mV, in 0.5 mV steps
 Accuracy: ± 5 %
Transvenous Pacer Pulse Simulation
Widths  Range: 0.1, 0.2, 0.5, 1, and 2 ms.
 Accuracy: ± 5 % of setting
Amplitudes  Range: 0 (off) and ± 2, ± 4, ±6, ±8, ±10, ±12. ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, ±700 mV
 Accuracy: ± (10 % setting + 0.2 mV)
Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent)  Lead II reference:
 Lead I – 67 %
 Lead II – 100 %
 Lead III – 33 %
 Lead V1 – 67%
 Lead V2 – 67%
 Lead V3 – 67 %
 Lead V4 – 67 %
 Lead V5 – 67 %
 Lead V6 – 67 %
 Lead I reference:
 Lead I – 100 %
 Lead II – 150 %
 Lead III – 50 %
 Lead V1 – 100%
 Lead V2 – 100%
 Lead V3 – 100 %
 Lead V4 – 100 %
 Lead V5 – 100 %
 Lead V6 – 100 %
Arrhythmia Selections
Pacer Interactive (Transcutaneous Pacer, Impulse 7000DP only)  Demand: 30 BPM to 360 BPM, in 1 BPM steps
Asynchronous
Non-Capture
Non-Function
Threshold (interactive pacing simulation only): 10 mA to 250 mA in 10 mA steps
Supraventricular  Atrial Fibrillation Coarse
Atrial Fibrillation Fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia
Paroxysmal Atrial Tachycardia
Nodal Rhythm
Supraventricular Tachycardia
Premature  Atrial PAC
Nodal PNC
PVC1 Left Ventricle (LV)
PVC1 Left Ventricle (LV) Early
PVC1 Left Ventricle (LV) R on T
PVC2 Right Ventricle (RV)
PVC2 Right Ventricle (RV) Early
PVC2 Right Ventricle (RV) R on T
Multifocal PVCs (LV and RV)
Ventricular  PVCs 6/min
PVCs 12/min
PVCs 24/min
Freq Multifocal
Trigeminy
Bigeminy
Pair PVCs
Run 5 PVCs
Run 11 PVCs
Monomorphic VTach: 120 to 300 (5 BPM steps)
Polymorphic VTach: 5 selections
Ventricular Fibrillation: Coarse and Fine
Asystole
Conduction  1° Block
2° Block Type I
2° Block Type II
3° Block
Right Bundle Branch Block (RBBB)
Left Bundle Branch Block (LBBB)
Transvenous Pacemaker (ECG simulation only)  Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
AV Sequential
Non-Capture
Non-Function
Selectable Pacer pulse Parameters for Transvenous Simulation (Atrial & Ventricular channels are independently selectable by the user)
Atrial Pacer Pulse (Note: relates only to the TV pacer)  Width: 0.1, 0.2, 0.5, 1, 2 ms
Polarity: + or –
Amplitude: 0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
Ventricular Pacer Pulse (Note: relates only to the TV pacer)  Width: 0.1, 0.2, 0.5, 1, 2 ms
Polarity: + or –
Amplitude: 0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
(BNC Jack)
Accuracy  ± 5 % for 2 Hz Square Wave , 50Ω output impedance

 

Models and Accessories

Model Name Description Price List
Impulse 6000D Impulse 6000D Defibrillator Analyzer 120 V – USAll models include:

  • USB Computer Communication Cable (cable assembly)
  • User Manual CD (CD-ROM)
  • Getting Started Guide (Manual)
  • Battery Eliminator (country specific)
  • Defib Paddle Contact Plates (FBC-IMP7000-8010)
  • Carrying Case (soft case)

 

Impulse 6000D-01 Impulse 6000D Schuko
Impulse 6000D-02 Impulse 6000D UK
Impulse 6000D-03 Impulse 6000D Japan
Impulse 6000D-04 Impulse 6000D Australia
Impulse 6000D-05 Impulse 6000D India
Impulse 6000D Defibrillator Analyzer Brazil Impulse 6000D Brazil

 

Combination Defibrillator and Pacer Adapters
I6/7 D/P ADPT101 GE Marquette defib/pacer adapter (part# 3065423)
(RESPONDER1500/1700 Series) (set of two) 4mm
I6/7 D/P ADPT104 Medtronic ERS/Physio-Control Defib/Pacer adapter (part# 3065461)
(QUIK COMBO) 4mm
I6/7 D/P ADPT108 Philips/Agilent Defib/Pacer adapter (part# 3065509)
HEARTSTART FR2/MRX 4mm
I6/7 D/P ADPT107 Philips/Agilent/HP Defib/Pacer adapter (part# 3065492)
(CODEMASTER Series-Round) 4mm
I6/7 D/P ADPT109 Zoll Medical PD2200 Multi-Function Series electrodes (part# 3065511)
M Series, M Series-CCT, R Series, E Series, AED Pro™ and AED Plus™ (Includes inline connector) (defib and pace) 4mm (AED Pro and AED Plus requires an auxiliary adapter part number: 8000-0804-01 available from Zoll Medical)

 

Defibrillator Adapters
FBC-IMP7000-8010 FBC-IMP7000-8010 defib plate (part# 3156262)
This item number is for a single plate assembly.
Each unit requires two.
I6/7 D/P ADPT102 Internal discharge paddle contacts (2) (4mm) (part# 3065438)
I6/7 D/P ADPT103 Kimberly Clark/R2 Darox MRL/MDE/NK: 4 mm defibrillator adapters (part# 3065450)
I6/7 D/P ADPT106 MedtronicERS/Physio-Control (part# 3065489)
(FAST PATCH) (set of two) 4mm
for Impulse 6000D/7000DP

 

Pacer Adapters
I6/7 D/P ADPT105 Medtronic ERS/Physio-Control pacer adapter (part# 3065477)
(QUIK PACE) (set of two): 4 mm
I6/7 D/P ADPT110 Zoll Medical NTP/PD1400:4 mm pacer adapters (part# 3065527)

 

Batteries, Chargers, and Power Supply
Fluke-744-8003 Fluke-744-8003 NIMH battery pack
(part# 665083) BP7235 rechargeable.
For Impulse 6000D & 7000DP

 

Software
Ansur IMPULSE 7000 Ansur Impulse 6000D/7000DP Plug-In license (part# 3091370)

 

Service Packages
Care Plans Request a quote

 

Related products

Fb-ps320_05b_c
Add to Wishlist
Quick View

Fluke Biomedical

PS320 Fetal Simulator

At week five, a fetus’ heart starts beating, increasing to 155 to 195 beats per minute prior to birth. The strength and number of beats can be measured by a fetal electronic monitor to determine whether a fetus is in distress. The PS320 fetal simulator mimics fetal and maternal heartbeats (ECG), along with uterine activity during labor to accurately test and troubleshoot fetal electronic monitors.

INCU II Incubator - Radiant Warmer Analyzerfb-incu-2_03a_ko_0
Add to Wishlist
Quick View

Fluke Biomedical

INCU II Incubator – Radiant Warmer Analyzer

Testing incubators and radiant warmers regularly is essential to validate their performance and ensure the safety and comfort of infants.
The Fluke Biomedical INCU II Incubator/Radiant Warmer Analyzer simplifies testing and verifying baby incubators, transport incubators, and radiant warmers are safe and conform to global IEC 60601-2-19 and IEC 60601-2-21 performance standards.

VT650 Gas Flow Analyzer
Add to Wishlist
Quick View

Fluke Biomedical

VT650 Gas Flow Analyzer

View multiple gas flow measurements simultaneously, with traceability to global regulatory standards.

ESA615 Electrical Safety Analyzer
Add to Wishlist
Quick View

Fluke Biomedical

ESA615 Electrical Safety Analyzer

The ESA615 Electrical Safety Analyzer brings fast and simple automated testing in the form of a portable analyzer to healthcare technology management professionals that perform electrical safety testing on medical equipment both in the field and in facilities.

ProSim 8 Vital Signs Patient Simulator
Add to Wishlist
Quick View

Fluke Biomedical

ProSim 8 Vital Signs Patient Simulator

Get the all-in-one multifunction vital signs simulator to ensure your devices provide the ultimate in patient safety.

IDA-1S Infusion Device Analyzer
Add to Wishlist
Quick View

Fluke Biomedical

IDA-1S Infusion Device Analyzer

Test a wide variety of infusion pumps quickly and accurately.

ProSim 2 and 3 Vital Signs Monitor Simulators
Add to Wishlist
Quick View

Fluke Biomedical

ProSim 2 and 3 Vital Signs Monitor Simulators

The ProSim 2 and ProSim 3 Vital Signs Simulators are the clear choice for biomedical engineering and field service technicians that need a quality feature rich device with high portability.

190M Medical ScopeMeter® Portable Oscilloscope
Add to Wishlist
Quick View

Fluke Biomedical

190M Medical ScopeMeter® Portable Oscilloscope

The 190M Medical ScopeMeter® Portable Oscilloscope is a high-performance, portable oscilloscope built upon the legacy of Fluke and Fluke Biomedical oscilloscopes in partnership with real customers like you.

Biomedical Test Equipment
graph-domotech

01 Biomedical

02 Diagnostic Imaging

03 Radiation Safety

04 Impact Indicators

05 Service & Calibration

Product Sites
  • Fluke Biomedical
  • Quantum Design
  • Montana Instruments
  • Hitachi-Aloka
  • IBP
  • Tristan Technology
  • MagnaBiosciences
  • IOG Products
Customer Care
  • About Us
  • Contact Us
  • Privacy Policy
  • Terms of Use
  • Sitemap
This website is proudly powered by the passionate team at Aleph IT!
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
  • Home
  • About Us
  • Contact Us
  • All Products
  • Fluke Biomedical
    • Defibrillator Analysers
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Neonatal Test Equipment
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Digital Pressure Meters
    • Patient Simulators
    • Phototherapy Radiometer
    • Test Automation
  • Quantum Design
  • Montana Instruments
  • Login
  • Newsletter

Login


Lost your password?