Skip to content
    • Contact
    • (02) 97486988
    • Contact
    • (02) 97486988
Domo-Technica AustraliaDomo-Technica Australia
  • About Us
  • Contact Us
  • 0
    Cart

    No products in the cart.

    Return to shop

  • Home
  • Fluke Biomedical
    • Defibrillator Analysers
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Neonatal Test Equipment
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Digital Pressure Meters
    • Patient Simulators
    • Phototherapy Radiometer
    • Rigid Endoscope Tester
    • Test Automation
  • Raysafe
    • X-Ray QA Instruments
    • Phantoms and Test Tools
    • Radiation Safety
    • Real-Time Personal Dosimeter
  • Quantum Design
    • Material Science
    • Cryogenics
    • Optics
    • Nanoscience
    • Sample Synthesis
    • Biotechnology & Chemistry
    • Magnetic Separator
    • Laboratory Squids
  • Montana Instruments
    • Research Platforms
    • Measurement Systems
    • Integrated Solutions
    • Accessories
  • Calibration
    • Australia
    • New Zealand
  • All Products
Add to Wishlist
SigmaPaceTM 1000 External Pacemaker Analyser
Home / Fluke Biomedical / Defibrillator Analysers

SigmaPaceTM 1000 External Pacemaker Analyser

The SigmaPace™ 1000 is a powerful handheld tool with a comprehensive range of test suites, measurement algorithms and test loads – as specified by the external pacemaker manufacturers – to fulfill your testing requirements both quickly and efficiently.

Categories: Defibrillator Analysers, Fluke Biomedical

Browse
  • Fluke Biomedical
    • Defibrillator Analysers
    • Digital Pressure Meters
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Neonatal Test Equipment
    • Patient Simulators
    • Phototherapy Radiometer
    • Rigid Endoscope Tester
    • Test Automation
  • Montana Instruments
    • Accessories
    • Integrated Solutions
    • Measurement Systems
    • Research Platforms
  • Quantum Design
    • Biotechnology & Chemistry
    • Cryogenics
    • Laboratory Squids
    • Magnetic Separator
    • Material Science
    • Nanoscience
    • Optics
    • Sample Synthesis
    • SCL-Sensor.Tech – Self-Sensing Cantilevers
  • Raysafe
    • Phantoms and Test Tools
    • Radiation Safety
    • Real-Time Personal Dosimeter
    • X-Ray QA Instruments
  • Uncategorised
  • Description
  • Features
  • Specifications
  • Models and Accessories
  • Video

SigmaPace 1000 External Pacemaker Analyzer

The SigmaPace™ 1000 is a powerful handheld tool with a comprehensive range of test suites, measurement algorithms and test loads – as specified by the external pacemaker manufacturers – to fulfill your testing requirements both quickly and efficiently.

*not available in Europe

 

 

 

Features

Standard Features
Unique full-featured biomedical test product
Tests for both transcutaneous and transvenous external pacemakers
Full range of user-selectable measurement algorithms and test loads for external pacemakers
Dual-channel signal acquisition for capturing synchronous AV-sequential pulse data
Interactive pacemaker and ECG simulation with 5-lead output
Large 21-character x 8-line alphanumeric LCD readout
Exclusive readout “HOLD” function
Exclusive SigmaPaceTM 1000 test features: DC static/dynamic leakage, and current drain

Specifications

 

Specifications
Modes of Operation
Manual
Remote (via standard RS232 serial port)
User Interface
Display: LCD readout
21 characters X 8 lines
Brightness/viewing angle adjustment
Keys: Eight push buttons (F-2, F-3 [UP arrow], F-4 [UP arrow], two DOWN arrows, ESCAPE, and ENTER)
Transcutaneous Pacer Tests
Output Pulse Measurement
Current: Ranges: 4.00 mA to 9.99 mA; 10.0 mA to 99.9 mA; 100 mA to 250 mA
Accuracy: ± 2 % of reading or ± 50 µA (whichever is greater)
Rate: Ranges: 5.0 PPM to 99.9 PPM; 100 PPM to 300 PPM
Accuracy: ± 0.5 % of reading or ± 0.3 PPM (whichever is greater)
Width: Ranges: 1.00 mS to 9.99 mS; 10.0 mS to 99.9 mS
Accuracy: ± 0.5 % of reading or ± 14 µS (whichever is greater)
Energy: Ranges: 1 µJ to 999 µJ; 1 mJ to 999 mJ; 1.00 J to 1.99 J
Accuracy: 5 % of reading/computation
Demand and Asynchronous Mode Tests
Waveform (Physiological Simulation): Normal sinus rhythm (NSR)
Complete P-QRS-T complex
Amplitude: 1.0mVpeak (lead I)
Modes of Operation: Underdrive: NSR @ 85 % of pulse interval/rate
Overdrive: NSR @ 115 % of pulse interval/rate
Auxiliary Control: NSR adjustable in 1-BPM increments
Auxiliary Rate Range: Underdrive 10 BPM (min); overdrive 300 BPM (max)
Active Outputs: 5-lead ECG; ventricular test load; high-level ECG jack
Pacemaker Compatibility: Pulse Rates: 30 to 200 PPM
Intended Types: Demand: VVI (pace and sense); async: VOO (pace)
Amplitude Sensitivity Test
Selections: R, S and T Waves Rate: 30 to 200 PPM
Test Loads: (30) 50 ? to 1550 ? in 50-? steps
Waveforms: Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ)
Amplitude: Range: 0.05 mVpeak to 5.0 mVpeak
Accuracy: ± 5 % of setting
Resolution: 0.05-mV steps (0.05 to 0.95 mVpeak); 0.50-mV steps (1 to 5 mVpeak)
Width: Range: 0.15 mS to 300 mS
Accuracy: ± 5 % of setting
Selections: 50
Resolution: 0.05-mS steps (0.15 to 0.95 mS); 1-mS steps (1 to 19 mS);
5-mS steps (20 to 95 mS) 25-mS steps (100 to 300 mS)
Active Outputs: 5-lead ECG; ventricular test load; high-level ECG jack
Pacemaker Compatibility: Pulse Rates: 30 to 200 PPM
Intended Type: VVI (pace and sense)
Noise Immunity/Line Frequency Test
Waveform: Sine wave
Frequency: 50 and 60 Hz
Accuracy: 0.5 Hz
Amplitude Testload Output: Range: 0 (OFF) to 10 mVpeak-to-peak
Accuracy: ± 5 % of setting
Resolution: 0.5-mVpeak-to-peak steps
Settings: (30) 50 ? to 1550 ? ± 1 %
5-lead ECG Output: Range: 0 (OFF) to 10 mVpeak-to-peak
Accuracy: ± 5 % of setting
Resolution: 0.50-mV steps
Reference: Lead I (RA to LA)
Active Outputs: 5-lead ECG; ventricular test load
Paced Refractory Period Test (PRP)
Range: 20 mS to 500 mS
Accuracy: 5 % of reading or 1 mS (whichever is greater)
Physiological Simulation: Selection: Single pulse, R Wave, SSQ
Pulse Width: 40 mS
Outputs: 5-lead ECG; ventricular test load
Pacemaker Compatibility: Pulse Rates: 30 to 200 BPM
Intended Type: VVI (pace and sense)
Sensed Refractory Period Test (SRP)
Range: 15 mS to 500 mS
Accuracy: ± 5 % of reading or ± 1 mS (whichever is greater)
Physiological Simulation: Selection: Double pulse, R Wave, SSQ
Pulse Width: 40 mS
Amplitude: 1 mVpeak lead I
Active Outputs: 5-lead ECG; ventricular test load
Pacemaker Compatibility: Pulse Rates: 30 to 200 BPM
Intended Type: VVI (pace and sense)
Test Loads
Transcutaneous Pacer: Selections: (31) 50 ? to 1550 ? in 50-? steps
Accuracy: ± 1 % of selection
Power Rating: 5 W (average); 40 W (peak) @ 1000 ?
Input Defibrillation Protection: Type: Internal spark gap
Episode Limit: 5 pulses @ 360 J (10 seconds min between discharges)
Life Limit: 250 pulses @ 360 J
Transvenous Pacer Tests
Output Pulse Measurement
Current: Ranges: 0.05 mA to .999 mA (available single channel only); 1.00 mA to 9.99 mA; 10 mA to 30 mA
Accuracy: ± 2 % of reading or ± 50 µA (whichever is greater)
Polarity Indicator: + or –
Rate: Ranges: 10.0 PPM to 99.9 PPM; 100 PPM to 999 PPM
Accuracy: ± 0.5 % or 0.3 PPM (whichever is greater)
Width: Ranges: 0.020 mS to .999 mS; 1.00 mS to 9.99 mS; 10.0 mS to 99.9 mS
Accuracy: 0.5 % or ± 14 µS (whichever is greater)
Resolution: ± 1 LSD or ± 4 µS (whichever is greater)
Voltage: Ranges: (available single channel only) 0.050 Vpeak to .999 Vpeak; 1.00 Vpeak to 9.99 Vpeak; 10 Vpeak to 30 Vpeak
Accuracy: ± 2 % of reading or ± 0.05 Vpeak (whichever is greater)
Polarity Indicator: + or –
Energy: Ranges: (available single channel only) 1 nJ to 999 nJ; 1 µJ to 999 µJ
Accuracy: ± 5 % of reading/computation
Display Formats: Atrial channel only; ventricular channel only; both A + V channels
AV Interval (Delay Time)
Measurement Ranges: 10.0 mS to 99.9 mS; 100 mS to 999 mS
Start Point: Atrial pulse leading edge
Stop Point: Ventricular pulse leading edge
Accuracy: 1 % of reading/computation
Demand/Async Mode Tests
Channels: Single and dual
Waveform: Sine square (SSQ)
Atrial Output: Simulated P Wave
Width: 30 mS
Amplitude: 2.0 mVpeak
Vent Output: Simulated R Wave
Width: 40 mS
Amplitude: 2.5 mVpeak
AV Interval: 90 mS (fixed)
Interactive Simulated Rates: Default Settings: Underdrive = NSR @ 85 % of pulse interval/rate; overdrive = NSR @ 115 % of pulse interval/rate
Manual: NSR simulations adjustable in 1-BPM increments
Limits: Underdrive (min) = 10 BPM; overdrive (max) = 300 BPM
Output: Ventricular channel test load; atrial channel test load
Pulse Rate: 30 to 200 PPM
Intended Pacemaker Types: Demand: VVI (V-channel pace and sense); AAI (A-channel pace and sense);
DDD (dual-channel pace and sense) Async/Continuous: VOO (V-channel pace and sense); AOO (A-channel pace and sense); DOO (dual-channel pace and sense)
Amplitude Sensitivity Test
Operation: Single-channel operation only (atrial or ventricular)
Atrial Channel (Physiological Simulation): Selection: P Wave
Rate: 30 to 120 BPM
Timing: Waveform delayed by 80 % of the pulse-to-pulse interval or 400 mS (whichever is shorter)
Active Output: Atrial test load
Available Test Loads: 200 ?, 500 ? (default setting) and 1000 ? ± 1 %
Waveform Selections: Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ) (default setting); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time
Sensitivity Waveform Amplitude:
– Test Load Selection: 500 ? (default setting)
— Range: 0.05 mVpeak to 50.0 mVpeak
— Accuracy: ± 5 % of setting
–Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1 to 50 mVpeak)
– Test Load Selection: 200 ?
— Range: 0.05 mVpeak to 20.0 mVpeak
— Accuracy: ± 5 % of setting
— Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1 to 20 mVpeak)
– Test Load Selection: 1000 ?
— Range: 0.05 mVpeak to 100 mVpeak-to-peak
— Accuracy: ± 5 % of setting
— Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1.0 to 49.5 mVpeak); 05.0 mVpeak (50 to 100 mVpeak)
Default Setting: 2.0 mVpeak
Widths: Range: 0.15 mS to 95.0 mS
Accuracy: ± 5 % of setting
Resolution: 0.05 mS (0.15 mS to 0.95 mS); 1 mS (1 mS to 19 mS); 5 mS (20 mS to 95 mS)
Intended Pacemaker Types: AAI (atrial pace and sense)
Pulse Rates: 30 to 200 PPM
Ventricular Channel (Physiological Simulation): Selection: R Wave, S Wave and T Wave
Rate: 30 to 120 BPM
Timing: Waveform delayed from the ventricular demand pacemaker pulse by 80 % of the pulse-to-pulse interval or 400 mS (whichever is shorter)
Active Output: Selected ventricular test load
Waveform Selections: Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ) (default setting); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time
Available Test Load(s): 200 ?, 500 ? (default setting) and 1000 ? ± 1 %
Amplitude:
– Pacer Load Selection: 500 ?
— Range: 0.05 to 50.0 mVpeak
— Accuracy: ± 5 % of setting
— Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1 to 50 mVpeak)
– Pacer Load Selection: 200 ?
— Range: 0.05 mVpeak to 20.0 mVpeak
— Accuracy: ± 5 % of setting
— Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1 to 20 mVpeak)
– Pacer Load Selection: 1000 ?
— Range: 0.05 mVpeak to 100 mVpeak-to-peak
— Accuracy: ± 5 % of setting
— Resolution: 0.05 mVpeak (0.05 to 0. 95 mVpeak); 0.50 mVpeak (1.0 to 49.5 mVpeak); 05.0 mVpeak (50 to 100 mVpeak)
Default Setting: 2.5 mVpeak
Widths: Range: 0.15 mS to 300 mS
Accuracy: ± 5 % of setting
Resolution: 0.05 mS (0.15 to 0.95 mS); 1 mS (1 to 19 mS); 5 mS (20 mS to 95 mS); 25 mS (100 mS to 300 mS)
Intended Pacemaker Type(s): VVI (atrial pace and sense only)
Pulse Rates: 30 to 200 PPM
Noise Immunity Test
Channels: Single, atrial or ventricular only
Waveform: Sine wave
Frequency: 50 and 60 Hz
Accuracy: ± 0.5 Hz
Active Output(s): Atrial- or ventricular-channel test load
Output Selections: Atrial channel only; ventricular channel only
ECG Signal: ECG signal can be added to the selected channel.
Atrial: 1 mVpeak
Ventricular: 1.5 mVpeak
Noise Line Amplitude:
– Pacer Load Selection: 500 ?
— Range: 0 (OFF) to 100 mVpeak-to-peak
— Accuracy: ± 5 % of setting
— Resolution: 5-mVpeak-to-peak steps
– Pacer Load Selection: 200 ?
— Range: 0 (OFF) to 40 mVpeak-to-peak
— Accuracy: ± 5 % of setting
— Resolution: 5-mVpeak-to-peak steps
– Pacer Load Selection: 1000 ?
— Range: 0 (OFF) to 200 mVpeak-to-peak
— Accuracy: ± 5 % of setting
— Resolution: 5-mVpeak-to-peak steps
Refractory Period Test (Atrial Channel)
Test Selections: Paced refractory period; sensed refractory period
Period: 20 to 500 mS
Accuracy: ± 5 % of reading (or ± 1 mS, whichever is greater)
Resolution: ± 1 LSD
Physiological Simulation: Selection: Square wave (default setting)
Atrial Channel: Simulated P Wave
Width: 1 mS
Amplitude: 20 mVpeak
Active Outputs: Atrial channel (4mm banana jacks) only
Additional Waveform Selections: Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time
Amplitude: Range: .05 mVpeak to 50.0 mVpeak
Accuracy: ± 5 % of setting
Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1.0 to 49.5 mVpeak )
Width: Range: 0.15 mS to 95.0 mS
Accuracy: ± 5 % of setting
Resolution: 0.05 mS (0.15 to 0.95 mS); 1 mS (1 to 19 mS); 5 mS (20 to 95 mS)
Active Outputs: Atrial channel (4mm banana jacks) only
Intended Pacemaker Types: AAI (atrial pace and sense only)
Pacemaker Rates: 30 to 200 PPM
Available Test Load: 500 ? ± 1 %
Refractory Period Test (Ventricular Channel)
Test Selections: Paced refractory period; sensed refractory period
Period: 20 to 500 mS
Accuracy: ± 5 % of reading (or ± 1 mS, whichever is greater)
Resolution: ± 1 LSD
Display Format: 3 digits
Physiological Simulation: Selection: Square wave (default setting)
Ventricular Channel: Simulated R Wave
Width: 1 mS
Amplitude: 20 mVpeak
Active Outputs: Ventricular channel (4mm banana jacks) only
Additional Waveform Selections: Square (SQU); triangle (TRI); haversine (HSN); sine square (SSQ); asymmetrical triangle (ISO) – fixed width: 2 mS rise time/13 mS fall time
Amplitude:
– Pacer Load Selection: 500 ?
— Range: .05 mVpeak to 50.0 mVpeak
— Accuracy: ± 5 % of setting
— Resolution: 0.05 mVpeak (0.05 to 0.95 mVpeak); 0.50 mVpeak (1.0 to 49.5 mVpeak)
– Default setting: 20 mVpeak
Width: Range: 0.15 mS to 300.0 mS
Accuracy: ± 5 % of setting
Resolution: 0.05 mS (0.15 to 0.95 mS); 1 mS (1 to 19 mS); 5 mS (20 to 95 mS); 25 mS (100 to 300 mS)
Default setting: 30 mS
Intended Pacemaker Types: VVI
Pacemaker Rates: 20 to 200 PPM
DC Leakage Current
Measurement Range: 00.1 µA to 99.9 µA
Input Polarity: Positive and negative
Resolution: 1 LSD (0.1 µA)
Display Format: 3 digits
Test Selections: Static: Continuous (power OFF)
Dynamic: Gated (power ON)
Test Load/Input Configurations: Atrial+ and atrial-
Ventricular+ and ventricular–
Atrial+ and ventricular+
Baseline/Test Selection: 500 ?
Dynamic Test Gating Algorithm: Measurement made 400 mS prior to the pacemaker pulse leading edge; 16 measurements averaged at a 4 mS rate for a total of 64 mS
Specified Applied Pacemaker Rate: 80 PPM
Current Drain Test
DC Current Ranges: 0.100 mA to 0.999 mA; 1.00 mA to 9.99 mA; 10.0 mA to 99.9 mA
Polarity: Positive or negative
Indicator: + or – symbol
Resolution: ± 1 LSD
Display Format: 3 digits plus decimal point
Accuracy: ± 5 % of reading ± 10 µA
Input DC Voltage: Nominal: ± 9 V
Range: 5.0 V to 10.5 V
Input Protection: Short-circuit protection
Protection Type: Internal in-line fast-acting 1/2 A fuse
Selectable Testloads: 200 ?, 500 ?, and 1000 ?
Battery Test Fixture: 9 V battery supply included, to facilitate connection of analyzer to recessed battery terminals within Medtronic 5388 and 5348 Temporary Pacemakers
Test Loads
Atrial Channel: Selections: 200 ?, 500 ?, and 1000 ?
Accuracy: ± 1 % of selection
Power Rating: 2 W
Ventricular Channel: Selections: 200 ?, 500 ?, and 1000 ?
Accuracy: ± 1 % of selection
Power Rating: 2 W
Tracking: Identical atrial and ventricular channel settings
Input Defibrillation Protection: Type: Internal spark gap
Episode Limit: 5 pulses @ 360 J (10 seconds min between discharges)
Life Limit: 250 pulses @ 360 J
Long-Term Test
Test Configuration: Transvenous Pacer: Atrial or ventricular channel only
Transcutaneous Pacer: Ventricular channel
Pulse Count Range: 999,999 (max)
Rate: 2 % to 20 % (default setting, 10 %)
Amplitude: 2 % to 20 % (default setting, 10 %)
Test Time (max): 999:59:59 (hhh:mm:ss)
Maximum Error Count: 200
Test Termination: Manual; or upon max error count
Testloads: 200 ?, 500 ?, and 1000 ?
Interactive Pacer ECG Simulation
Simulation of demand, continuous, noncapture, and nonfunction patient -ECG activity
Additional user
-selectable parameters: NSR Heart Rate: Asystole and 20 to 250 BPM (1-BPM steps)
NSR PR Interval: 0.05 to 0.30 s (6 settings)
Pacemaker Capture/Threshold: Transcutaneous: 10 to 250 mA (10-mA steps)
Transvenous: 1 to 25 mA (1-mA steps)
General Specifications
Serial Port: Type: RS232
Connector Type: DB-9 (male)
Baud Rates: 2400, 9600, and 19200
Data Control: Xon/Xoff
Power Requirements: External battery charger source/power supply
100 to 240 Vac, 50/60 Hz operation
Auto power-off feature during battery operation
Battery Life: 20 hours (min)
Environmental Specifications
Temperature Range: Operating: 15 °C to 35 °C (59 °F to 95 °F)
Storage: 0 °C to 50 °C (32 °F to 122 °F)
Humidity Range: < 90 % noncondensing
Dimensions
Size: 8 L X 4 W X 2 H (approx) (203 mm L X 101 mm W X 50 mm H)
Weight: 2 lb (approx) (0.90 kg)
CE Mark
User Safety:
EMC: EN61326-1.1997
Conforms to: UL STD 3101-1
Certified to: CAN/USA STD C22.2 No.1010

 

Models and Accessories

Model Name Description Price List
SIGMAP1K-USA120V Sigmapace 1000 External Pacemaker Analyzer Tests both transcutaneous & transvenous external pacemakers, USA, 120 V (Item #2247700)

Includes:

  • Operator’s manual (Item #2243306)
  • Nylon carrying case (Item #2392906)
  • Transvenous pacer test leads 2 sets (2 each red leads (Item #2201166), and black leads (Item #2201153))
  • SigmaPace 9 V dc load test cable (Item #2392272)
  • Serial interface cable (Item #2392260)
  • Universal-input battery charger (Item #2184298)
  • Power cord set USA 120 V ac

 

SIGMAP1K-JPN100V Sigmapace 1000, Japan, 100 V (Item #2394548)

Includes:

  • Operator’s manual (Item #2243306)
  • Nylon carrying case (Item #2392906)
  • Transvenous pacer test leads, 2 sets (2 each red leads (Item #2201166), and black leads (Item #2201153))
  • SigmaPace 9 V dc load test cable (Item #2392272)
  • Serial interface cable (Item #2392260)
  • Universal-input battery charger (Item #2184298)
  • Power cord set USA 120 V ac

 

SIGMAP1K-SHK250V Sigmapace 1000, Schuko, 250 V (Item #2394553)

Includes:

  • Operator’s manual (Item #2243306)
  • Nylon carrying case (Item #2392906)
  • Transvenous pacer test leads, 2 sets (2 each red leads (Item #2201166), and black leads (Item #2201153))
  • SigmaPace 9 V dc load test cable (Item #2392272)
  • Serial interface cable (Item #2392260)
  • Universal-input battery charger (Item #2184298)
  • Power cord set USA 120 V ac

 

SIGMAP1K-UK250V Sigmapace 1000, United Kingdom, 250 V (Item #2394566)
Includes:

  • Operator’s manual (Item #2243306)
  • Nylon carrying case (Item #2392906)
  • Transvenous pacer test leads, 2 sets (2 each red leads (Item #2201166), and black leads (Item #2201153))
  • SigmaPace 9 V dc load test cable (Item #2392272)
  • Serial interface cable (Item #2392260)
  • Universal-input battery charger (Item #2184298)
  • Power cord set USA 120 V ac

 

 

Test Leads, Probes and Clips
3010-0610 3010-0610 SGMA TV test lead – black
(part# 2201153) for SigmaPace 1000
Used for testing invasive pacemakers. Two (2) required for dual chamber testing.
3010-0611 3010-0611 SGMA TV test lead – red
(part# 2201166) for SigmaPace 1000
Used for testing invasive pacemakers. Two (2) required for dual chamber testing.

 

Combination Defibrillator and Pacer Adapters
3010-0609 3010-0609, Zoll PD & M-Series adapter (part# 2201148)
(for use with SigmaPace 1000)

 

Pacer Adapters
3010-0603 3010-0603 Physio Quik Pace (part# 2201088)
(for use with SigmaPace 1000)
3010-0604 3010-0604 Physio Quik Combo (part# 2201095)
(for use with SigmaPace 1000)
3010-0605 3010-0605 Agilent HP CodeMaster adapter (part# 2201109)
(for use with SigmaPace 1000)
3010-0606 3010-0606 MQ Responder (part# 2201111)
GE Marquette Medical electrode adapter (for use with SigmaPace 1000)
3010-0608 3010-0608, Zoll NTP adapter (part# 2201130)
(for use with SigmaPace 1000)

 

Batteries, Chargers, and Power Supply
3010-0655 Detachable AC power cord set (UK) (part# 2201428)
IEC 320 C6 type 3-pin inlet (for SigmaPace 1000)
3010-0654 IEC 320 C6 type 3-pin AC power cord set (part# 2201419)
for ProSim 8 or SigmaPace 1000 (Japan)
Sigma1K-8001 Sigma1K-8001 SigmaPace 1000 battery pack
(part# 2229830)

 

Cables
3010-0585FG Serial Interface Cable, DB9F to DB9F (PC to SigmaPace™1000) (part#2392260) (this requires F-M gender changer)
3010-0602FG 3010-0602FG, 9VDC Load Test Cable (part# 2392272)
(for use with SigmaPace 1000)
3010-0639 3010-0639 Philips Agilent Cable (part# 2201323)
(for use with SigmaPace 1000)

 

Service Packages
Care Plans Request a quote

 

Video

Related products

Impulse 7010 Defibrillator Selectable Load Accessory
Add to Wishlist
Quick View

Fluke Biomedical

Impulse 7010 Defibrillator Selectable Load Accessory

As you know the Impulse 7000 can do the following defibrillator performance tests: ECG functionality tests, AED or Shock Advisory capability tests, Pacemaker tests, and charge- and synch-time tests.

ProSim 4 Vital Signs Patient Simulator
Add to Wishlist
Quick View

Fluke Biomedical

ProSim 4 Vital Signs Patient Simulator

ProSim 4 Vital Signs Simulator with breakthrough touchscreen technology offers quick and simple one-tap testing for patient monitor performance checks and troubleshooting

ProSim SPOT Light SpO2 Pulse Oximeter Analyzer
Add to Wishlist
Quick View

Fluke Biomedical

ProSim SPOT Light SpO2 Pulse Oximeter Analyzer

Featuring an exclusive ergonomic design, the ProSim SPOT Light is the first comprehensive SpO2 functional tester to come in a handheld and easy-to-use device.

ProSim 2 and 3 Vital Signs Monitor Simulators
Add to Wishlist
Quick View

Fluke Biomedical

ProSim 2 and 3 Vital Signs Monitor Simulators

The ProSim 2 and ProSim 3 Vital Signs Simulators are the clear choice for biomedical engineering and field service technicians that need a quality feature rich device with high portability.

IDA-1S Infusion Device Analyzer
Add to Wishlist
Quick View

Fluke Biomedical

IDA-1S Infusion Device Analyzer

Test a wide variety of infusion pumps quickly and accurately.

ProSim 8 Vital Signs Patient Simulator
Add to Wishlist
Quick View

Fluke Biomedical

ProSim 8 Vital Signs Patient Simulator

Get the all-in-one multifunction vital signs simulator to ensure your devices provide the ultimate in patient safety.

ESA615 Electrical Safety Analyzer
Add to Wishlist
Quick View

Fluke Biomedical

ESA615 Electrical Safety Analyzer

The ESA615 brings fast and simple automated testing to a portable analyzer.

DPM4 Parameter Testers
Add to Wishlist
Quick View

Fluke Biomedical

DPM4 Parameter Testers

DPM4 Parameter Tester is a highly accurate meter for testing a wide range of medical devices. Key features include its lightweight, compact size and battery operation

Biomedical Test Equipment
graph-domotech

01 Biomedical

02 Diagnostic Imaging

03 Radiation Safety

04 Impact Indicators

05 Service & Calibration

Product Sites
  • Fluke Biomedical
  • Quantum Design
  • Montana Instruments
  • Hitachi-Aloka
  • IBP
  • Tristan Technology
  • MagnaBiosciences
  • IOG Products
Customer Care
  • About Us
  • Contact Us
  • Privacy Policy
  • Terms of Use
  • Sitemap
This website is proudly powered by the passionate team at Aleph IT!
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
  • Home
  • About Us
  • Contact Us
  • All Products
  • Fluke Biomedical
    • Defibrillator Analysers
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Neonatal Test Equipment
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Digital Pressure Meters
    • Patient Simulators
    • Phototherapy Radiometer
    • Test Automation
  • Quantum Design
  • Montana Instruments
  • Login
  • Newsletter

Login


Lost your password?