Skip to content
    • Contact
    • (02) 97486988
    • Contact
    • (02) 97486988
Domo-Technica AustraliaDomo-Technica Australia
  • About Us
  • Contact Us
  • 0
    Cart

    No products in the cart.

    Return to shop

  • Home
  • Fluke Biomedical
    • Defibrillator Analysers
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Neonatal Test Equipment
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Digital Pressure Meters
    • Patient Simulators
    • Phototherapy Radiometer
    • Rigid Endoscope Tester
    • Test Automation
  • Raysafe
    • X-Ray QA Instruments
    • Phantoms and Test Tools
    • Radiation Safety
    • Real-Time Personal Dosimeter
  • Quantum Design
    • Material Science
    • Cryogenics
    • Optics
    • Nanoscience
    • Sample Synthesis
    • Biotechnology & Chemistry
    • Magnetic Separator
    • Laboratory Squids
  • Montana Instruments
    • Research Platforms
    • Measurement Systems
    • Integrated Solutions
    • Accessories
  • Calibration
    • Australia
    • New Zealand
  • All Products
Add to Wishlist
RaySafe X2 X-ray measurement system
Home / Raysafe / X-Ray QA Instruments

RaySafe X2 X-ray measurement system

Your x-ray dosimeter solution, providing x-ray test and measurement sensors for R/F, MAM, CT, Survey, and light applications, including X2 View software.

Categories: Raysafe, X-Ray QA Instruments

Browse
  • Fluke Biomedical
    • Defibrillator Analysers
    • Digital Pressure Meters
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Neonatal Test Equipment
    • Patient Simulators
    • Phototherapy Radiometer
    • Rigid Endoscope Tester
    • Test Automation
  • Montana Instruments
    • Accessories
    • Integrated Solutions
    • Measurement Systems
    • Research Platforms
  • Quantum Design
    • Biotechnology & Chemistry
    • Cryogenics
    • Laboratory Squids
    • Magnetic Separator
    • Material Science
    • Nanoscience
    • Optics
    • Sample Synthesis
    • SCL-Sensor.Tech – Self-Sensing Cantilevers
  • Raysafe
    • Phantoms and Test Tools
    • Radiation Safety
    • Real-Time Personal Dosimeter
    • X-Ray QA Instruments
  • Uncategorised
  • Description
  • Features
  • Specification
  • Models and Accessories
  • Video

X-ray dosimetry monitoring at your fingertips
We’ve grown accustomed to intuitive interactions with our devices. After all, it’s not the device that’s most important, but what you can achieve with it that matters. Simplicity has always been a hallmark for RaySafe. But simplicity on the outside requires a lot of work on the inside. Our simplicity comes from a careful balance of advanced sensor technology paired with electronic wizardry and intelligent signal processing. The result – a radiation dosimeter system that provides accurate X-ray measurements with the ultimate in user friendliness. Right at your fingertips.

Less effort. More insight.
Life is busier. There are more demands on your time. So you need to remove the unnecessary steps in taking X-ray measurements. Like positioning the sensor, choosing a setting, or interpreting the results. Fortunately, the X2 R/F sensor is orientation independent so the only thing you need to do is to place it in the X-ray beam and turn on the instrument. The rest is automatic – no menus, no selections.

Full range x-ray measurements
Ease-of-use means you get everything you need in one exposure, with one sensor – automatically. The RaySafe X2 offers X-ray measurement sensors for R/F, MAM, CT, Survey, and even light applications. Choose the sensors you need and add the ones you will need at a later stage. The X2 X-ray sensors are made without the need to select ranges or special modes. Most sensors also measure waveforms that can be analyzed directly on the base unit.

Intuitive interface and first-class precision
The RaySafe X2 X-ray measurement sensors and electronics are specifically designed to minimize the need for user interaction. A groundbreaking concept in sensor design and circuitry provides unsurpassed accuracy, reproducibility and sensitivity. Intelligent algorithms clearly indicate when a parameter is outside its specified range.

Finally, a built-in, self-test system ensures your system is in complete working order. This provides added peace of mind and further assures accurate measurements the first time and every time.

Features

All exposure data are saved in the X2 Base Unit. In each session, you can swipe to view measurement data from previous exposures for reference or comparison. A full session of measurements can easily be uploaded onto a PC at a later stage with help of the X2 View software.

The RaySafe X2 touch screen interface allows the user to view data in a comprehensive yet flexible way. The home screen displays every available parameter from the attached sensor. To zoom in on any parameter, just tap it for a larger view, and pinch to zoom into details. Use a quick swipe to display waveforms. Navigation is made simple by using common Menu, Home and Back keys. All exposures are saved in the base unit. In each session, you can swipe to quickly go back to previous exposures for reference or comparisons. A full session of X-ray measurements can be uploaded to the X2 View software at a later stage for more manipulation.

Specification

X2 General Specifications

EMC According to IEC 61326-1
Safety According to IEC 61010-1
X-Ray Meters Standard Complies with IEC 61674
Exposures Needed One
USB Cables 2 m (6.6 ft), 5 m (16.4 ft) and 5 m active extender
Size of Base Unit 34 x 85 x 154 mm (1.3 x 3.3 x 6.1 in)
Weight of Base Unit 521 g (18.4 oz)
Operating Temperature 15 – 35 °C (59 – 95 °F)
Storage temperature -25 – 70 °C (-13 – 158 °F)
Power source Rechargeable Li ion battery
Battery Time ~ 10 hours intensive usage
Battery tested According to UN 38.3
Display 4.3″ LCD with capacitive touch
Memory ~ 10000 latest exposures
Software X2 View for data handling and analysis. Also exports data to Microsoft Excel.
PTB Certificate DE-17-M-PTB-0053

 

X2 mAs
mAs

Range 0.001 – 9999 mAs
Resolution 0.001 mAs
Uncertainty 1 %
mA
Range (peak) 0.1 – 1500 mA
Resolution 0.01 mA
Uncertainty 1 %
Time
Range 1 ms – 999 s
Resolution 0.1 ms
Bandwidth 1 kHz
Uncertainty 0.5 %
Pulses
Range 1 – 9999 pulses
Resolution 1 pulse
Pulse rate
Range 0.1 – 200 pulses/s
Resolution 0.1 pulse/s
mAs/Pulse
Range 0.001 – 9999 mAs
Resolution 0.001 mAs
Uncertainty 1 %
Waveform
Resolution 125 μs*
Bandwidth 1 kHz

 

* automatically reduced for exposures longer than 3 s.

X2 R/F Sensor

Size of Detector 14 x 22 x 79 mm (0.5 x 0.9 x 3.1 in)
Weight 42 g (1.5 oz)
Active Compensation
Dose/dose rate 40 – 150 kVp, 1 – 14 mm Al HVL
kVp 40 – 150 kVp, up to 1 mm Cu
TF 60 – 120 kVp, up to 1 mm Cu
Dose
Range 1 nGy – 9999 Gy (0.1 μR – 9999 R)
Uncertainty 5 % or 5 nGy (0.5 μR)
Dose rate
Range 1 nGy/s – 500 mGy/s (5 μR/min – 3400 R/min)
Resolution 1 nGy/s (5 μR/min)
Trig level 50 nGy/s (340 μR/min)
Uncertainty 5 % or 10 nGy/s (70 μR/min) x duty cycle
kVp
Range 45 – 150 kVp
Minimum dose 50 μGy (6 mR)
Minimum dose rate (peak) 10 μGy/s (70 mR/min)
Uncertainty 2 %
HVL
Range 1 – 14 mm Al
Minimum dose 1 μGy (120 μR)
Minimum dose rate (peak) 0.5 μGy/s (3.5 mR/min) at > 70 kV
2.5 μGy/s (17 mR/min) at 50 kV
Uncertainty 10 %
Total filtration
Range 1.5 – 35 mm Al
Minimum dose 50 μGy (6 mR)
Minimum dose rate (peak) 10 μGy/s (70 mR/min)
Uncertainty 10 % or 0.3 mm Al
Time
Range 1 ms – 999 s
Resolution 0.1 ms
Bandwidth 4 Hz – 4 kHz* (*auto adjusted based on signal level)
Uncertainty 0.5 %
Pulses
Range 1 – 9999 pulses
Minimum dose rate (peak) 0.5 μGy/s (3.5 mR/min)
Pulse rate
Range 0.1 – 200 pulses/s
Minimum dose rate (peak) 0.5 μGy/s (3.5 mR/min )
Dose/pulse
Range 1 nGy/pulse – 999 Gy/pulse
(0.1 μR/pulse – 999 R/pulse)
Minimum dose rate (peak) 0.5 μGy/s (3.5 mR/min)
Waveforms
Resolution 62.5 μs*
Bandwidth kV 0.1 – 0.4 kHz**
Bandwidth dose rate 4 Hz – 4 kHz**

 

* automatically reduced for exposures longer than 1.5 s
** automatically adjusted depending on signal level

X2 MAM Sensor

Size detector 14 x 22 x 79 mm (0.5 x 0.9 x 3.1 in)
Weight 42 g (1.5 oz)
Active Compensation – Beam quality independent for the following ranges:

Dose/dose rate & HVL (No selections needed. With or without paddle or phantom.)

Mo/Mo, Mo/Rh 20 – 40 kVp
Rh/Ag 27 – 40 kVp
Mo/Al, W/Rh, W/Ag, W/Al, Rh/Rh, Rh/Al 20 – 50 kVp
Mo/Cu, Rh/Cu, W/Cu, W/Ti 40 – 50 kVp

kVp (User selectable beam qualities. Paddle compensation available when relevant.)

W/Ag 20 – 40 kVp
W/Al 20 – 50 kVp
Measuring over 40 kVp requires
an X2 R/F Sensor + 2 mm Al (incl.)
W/Rh 20 – 40 kVp
Mo/Mo 20 – 40 kVp
Mo/Rh 32 – 40 kVp using + 2 mm Al (incl)
Rh/Ag 27 – 40 kVp
Mo/Cu, W/Cu, W/Ti 40 – 50 kVp, using the X2 R/F Sensor
Dose
Range 1 μGy – 9999 Gy (0.1 mR – 9999 R)
Uncertainty 5 %
Dose Rate
Range 10 μGy/s – 300 mGy/s (70 mR/min – 2000 R/min)
Uncertainty 5 %
kVp
Range 20 – 50 kVp*
Measuring above 40 kVp
requires an X2 R/F Sensor
and on W/Al +2 mm Al (incl.)
(*depending on beam quality, see active compensation)
Minimum dose 50 μGy (6 mR)
Minimum dose rate (peak) 10 μGy/s (70 mR/min)
Uncertainty 2 % or 0.5 kV (without paddle)
2 % or 0.7 kV (with paddle)
HVL
Range 0.2 – 3.6 mm Al
Minimum dose 1 μGy (0.1 mR)
Uncertainty 5 % above 25 kV; 10% below 25 kV
Time
Range 1 ms – 999 s
Resolution 0.1 ms
Bandwidth 400 Hz
Uncertainty 0.5 %
Pulses
Range 1 – 9999 pulses
Pulse rate
Range 0.1 – 200 pulses/s
Dose/pulse
Range 1 μGy/pulse – 999 Gy/pulse
(0.1 mR/pulse – 999 R/pulse)
Waveforms
Resolution 62.5 μs*
Bandwidth 400 Hz

 

* automatically reduced for exposures longer than 1.5 s

X2 Light Sensor

Weight 136 g (4.8 oz)
Size 48 x 60 x 68 mm (1.9 x 2.4 x 2.7 in)
Classification DIN 5032 part 7 class B
Standards Complies with relevant parts of
AAPM TG18, IEC 62563-1 and
IEC 61223-2-5.
Luminance
Range 0.01 – 10 000 cd/m² (0.03 – 34 000 fL)
Resolution 0.001 cd/m² (0.001 fL)
Aperture angle 5°
Measurement area Ø 10 mm (0.4 in)
Uncertainty illuminant A 3%
Deviation from human eye V(λ) (f1’) < 3 % (see figure Photopic Response)
Illuminance
Range 0.1 – 100 000 lux (0.01 – 9000 fc)
Resolution 0.01 lux (0.001 fc)
Uncertainty illuminant A 3%
Deviation from human eye V(λ) (f1’) < 3 % (see figure Photopic Response)
Cosine deviation (f2) < 3 % (see figure Cosine Response)

 

X2 Survey Sensor

Dimensions 14 x 66 x 192 mm (0.5 x 2.6 x 7.6 in)
Weight 140 g (4.9 oz)
Active Compensation H*(10) – when selecting Sv
Air kerma – when selecting Gy or R
Trig Modes
Manual Manual start and stop of measurement
Auto Trig level (N80): 10 μGy/h (1.2 mR/h)
or 20 μSv/h
H*(10)
Range 0 nSv – 9999 Sv
Resolution 1 nSv
Uncertainty 10 %, N-series 20 – 150 kV
H*(10) Rate
Range 0 μSv/h – 150 mSv/h
Uncertainty 10 % or 0.3 μSv/h, N-series 20 – 150 kV
Air Kerma
Range 0 nGy – 9999 Gy (0 μR – 9999 R)
Resolution 1 nGy (0.1 μR)
Uncertainty 5 %, RQA 50 – 150 kV
10 %, N-series 40 – 150 kV
Air Kerma Rate
Range 0 μGy/h – 100 mGy/h (0 mR/h – 10 R/h)
Uncertainty 5 % or 0.3 μGy/h (0.03 mR/h),
RQA 50 – 150 kV
10% or 0.3 μGy/h (0.03 mR/h),
N-series 40 – 150 kV
Mean Energy
Range 30 – 120 keV
Uncertainty 10 %
Minimum Dose Rate 10 μSv/h or 10 μGy/h (1 mR/h)
Defining Standard ISO 4037-1
Time
Range 0.1 – 9999 s
Resolution 0.01 s
Bandwidth 1 Hz
Waveform
Range 10 ms
Bandwidth 1 Hz
Minimum Dose Rate 1 μSv/h or 1 μGy/h (0.1 mR/h)

 

X2 CT Sensor

Size 14 x 22 x 219 mm (0.5 x 0.9 x 8.6 in)
Weight 86 g (3.0 oz)
Size Ø 12.0 mm (0.47 in)
Standard For measurements in accordance with IEC 60601-2-44
Active length 100 mm (3.94 in)
Energy dependence < 5 % for 70 – 150 kV
(RQR, RQA and RQT beam qualities)
Automatic environmental compensation 55 – 110 kPa, 15 – 35 °C (59 – 95 °F)
Dose
Range 10 μGy – 999 Gy (1 mR – 999 R)
Uncertainty 5 %
Dose length product
Range 100 μGycm – 9999 Gycm (10 mRcm – 9999 Rcm)
Uncertainty 5 %
Dose Rate
Range 10 μGy/s – 250 mGy/s (70 mR/min – 1700 R/min)
Uncertainty 5 %
Time
Range 10 ms – 999 s
Resolution 1 ms
Bandwidth 10 Hz
Uncertainty 0.5 %
Waveforms
Resolution 1 ms
Bandwidth 10 Hz

 

Models and Accessories

Model Name Description Price List
1506013 RaySafe X2 Prestige – 1506013
Includes X2 Base unit with mAs, X2 View software, and the following:

  • X2 R/F Sensor
  • X2 MAM Sensor
  • X2 CT Sensor
  • X2 Light Sensor

 

1506014 RaySafe X2 Prestige – 1506014
Includes X2 Base unit, X2 View software, and the following:

  • X2 R/F Sensor
  • X2 MAM Sensor
  • X2 CT Sensor
  • X2 Light Sensor

 

1506035 RaySafe X2 Prestige -1506035
Includes X2 Base unit with mAs, X2 View software, and the following:

  • X2 R/F Sensor
  • X2 MAM Sensor
  • X2 CT Sensor
  • X2 Survey Sensor

 

1506036 RaySafe X2 Prestige -1506036
Includes X2 Base unit, X2 View software, and the following:

  • X2 R/F Sensor
  • X2 MAM Sensor
  • X2 CT Sensor
  • X2 Survey Sensor

 

1506037 RaySafe X2 Prestige -1506037
Includes X2 Base unit with mAs, X2 View software, and the following:

  • X2 R/F Sensor
  • X2 MAM Sensor
  • X2 Light Sensor
  • X2 Survey Sensor

 

1506038 RaySafe X2 Prestige -1506038
Includes X2 Base unit, X2 View software, and the following:

  • X2 R/F Sensor
  • X2 MAM Sensor
  • X2 Light Sensor
  • X2 Survey Sensor

 

 

Service Packages
Care Plans Request a quote

 

Video

Related products

Dual-Diode Dosimeter Patient Dose Monitor
Add to Wishlist
Quick View

Raysafe

Dual-Diode Dosimeter Patient Dose Monitor

Dual-Diode Dosimeter Patient Dose Monitor provides instant patient data on radiation exposure to sensitive organs and rapid checks of equipment output

07-443 Hand Held Deluxe Digital Clamshell Densitometer
Add to Wishlist
Quick View

Raysafe

07-443 Hand Held Deluxe Digital Clamshell Densitometer

Get all the benefits of state-of the-art features in a compact, hand held unit. Become a “speed reader with this accurate, rugged portable Clamshell Densitometer.

RaySafe P Fluoro Phantom
Add to Wishlist
Quick View

Raysafe

RaySafe P Fluoro Phantom

Tests and analysis of digital and analogue radiography and fluoroscopy equipment

451 Assistant for Excel
Add to Wishlist
Quick View

Fluke Biomedical

451 Assistant for Excel

The 451EXL provides remote control for many of the 451B and 451P functions via a Microsoft® Excel-based user interface, including real-time data logging with user defined alarm parameters, upload of the internal data log into Excel worksheet, real-time virtual instrument display, and accumulated dose measurement over a user defined integration period.

451B Ion Chamber Survey Meter with Beta Slide
Add to Wishlist
Quick View

Raysafe

451B Ion Chamber Survey Meter with Beta Slide

The auto-ranging 451B ion chamber survey meter measures radiation rate and accumulated dose from beta, gamma and x-ray radiation sources.

RaySafe P 3D Fluoro Phantom
Add to Wishlist
Quick View

Raysafe

RaySafe P 3D Fluoro Phantom

Test the cone beam functionality of fluoroscopic x-ray systems

RaySafe Pro-Slit Phantom
Add to Wishlist
Quick View

Raysafe

RaySafe Pro-Slit Phantom

Slit camera for focal spot measurement

489-50 Gamma Scintillation Probe 1
Add to Wishlist
Quick View

Fluke Biomedical

489-50 Gamma Scintillation Probe 1

The Geiger-Mueller (GM) detectors fulfill a wide variety of radiation measurement needs. Probes are available for detection of alpha, beta, and gamma radiation.

Biomedical Test Equipment
graph-domotech

01 Biomedical

02 Diagnostic Imaging

03 Radiation Safety

04 Impact Indicators

05 Service & Calibration

Product Sites
  • Fluke Biomedical
  • Quantum Design
  • Montana Instruments
  • Hitachi-Aloka
  • IBP
  • Tristan Technology
  • MagnaBiosciences
  • IOG Products
Customer Care
  • About Us
  • Contact Us
  • Privacy Policy
  • Terms of Use
  • Sitemap
This website is proudly powered by the passionate team at Aleph IT!
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
  • Home
  • About Us
  • Contact Us
  • All Products
  • Fluke Biomedical
    • Defibrillator Analysers
    • Electrical Safety Analysers
    • Electrosurgery Analysers
    • Gas Flow Analysers
    • Neonatal Test Equipment
    • Infusion Device Analysers
    • Medical ScopeMeter
    • Digital Pressure Meters
    • Patient Simulators
    • Phototherapy Radiometer
    • Test Automation
  • Quantum Design
  • Montana Instruments
  • Login
  • Newsletter

Login


Lost your password?